BCFDM_a078 — Door Unzeroed
Full alert identifier: BCFDM_a078_DoorUnzeroed
Alert description
The door zero position is cleared.
Potential impact
The door power operation is disabled.
Set condition
The door zero position was cleared due to latch closing while door angle is above threshold value, the angle reaching threshold negative value, or the door controller resetting while latch is not closed. Alert payload data should provide more information on specific condition(s) involved.
Clear condition
The door zero position is relearned.
Signals to check
These diagnostic fields are associated with BCFDM_a078. Check the alert's Payload in Service Mode Plus for available values.
| Signal | Unit |
|---|---|
Door Unzeroing Reason
BCFDM_
|
— |
Check other alerts
With Service Mode Plus enabled, open Service Alerts → All. Review the other alerts for clues to the cause of BCFDM_a078.